6 World Precision Instruments
INSTRUMENT DESCRIPTION
Secondary Vent Port
(Vacuum/Atmosphere)
econ
Solenoid Input
Selector Switch
Hold Pressure
Gauge
Hold Pressure
Gauge
Vacuum
Pressure
Gauge
Eject Pressure
Gauge
Eject Pressure
Gauge
PV820
PV830
Solenoid
Controls
Solenoid
Controls
Eject Port
noi
u
Power
Switch
Secondary Vent Port
Selector Switch
acuum
Pr
ssur
Gau
Fig. 7—The major features of the PicoPumps are labeled in this diagram.
Regulators/Gauges– The PV820 has a hold pressure regulator and an eject pressure
regulator. In addition to those two, the PV830 also has a vacuum regulator. Each
precision regulator is connected with a measuring gauge for monitoring the pressure that
fl ows through the regulator.
Adjust the pressure by turning the Regulator knobs located near the pressure gauges.
Turn the knob clockwise to increase the pressure, and counter-clockwise to decrease it.
Because the regulating mechanism is self-venting, pressure is automatically released
when the regulator setting is decreased. Although the pressure gauges cannot be read
with high accuracy (especially at pressure settings below 2PSI), reproducible pressures
may be obtained, because the pressure regulators have a 20-turn resolution.
The pressure regulating mechanism functions by continuously bleeding a small amount
of gas. This bleed rate increases as the difference between the input pressure and the
output pressure increases. Loss of gas may be minimized by decreasing the difference
between the input and output pressures, but some decrease in regulation may be noticed
if the input pressure is not at least 10% greater than the output pressure.
!
CAUTION: Be careful NOT apply too much pressure when turning the knobs fully
counterclockwise to minimum pressure. They cannot be adjusted to 0.0 PSI.
Over-tightening the knobs damages the regulators.