WPI PZMIV User Manual Page 10

  • Download
  • Add to my manuals
  • Print
  • Page
    / 14
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 9
229
MICROSCOPES, CAMERAS
Maintenance of the desired
temperature set-point can
now be achieved through the
use of a precision, shielded
temperature probe coupled to
a peripheral Air-Therm heating
unit.
Precision: Noise is a major
concern in any experimental
application. WPI's Air-Therm™
is electrically quiet as well
as acoustically quiet). Since
the Air-Therm can be placed
away from the incubator,
any residual electrical and
vibrational interference
from the heater is virtually
eliminated. Accuracy of ±0.1°C
can be routinely maintained at
the sample site (using a fully
programmable differential
algorithm to automatically
adjust the temperature on
the Air-Therm unit). Air-
Therm maintains a uniform
temperature across the
microscope stage (allowing for
uniform heating of multiwell
dishes). These tight tolerances
result in minimal focal drift
once sample/incubator equilibrium has been achieved.
Glass Bottom Dishes: For high numerical aperture (NA) live-cell
imaging, cells must be plated on glass (e.g., WPI’s FluoroDish Glass
Bottom Dishes, spotting 0.17 mm glass thickness) or standard cover slips.
Since cover slips inherit other problems associated with sterility and
accidental damage to expensive microscope condensers, WPI designed
new and improved glass bottom dishes (35 mm x 10 mm or 35 mm X
22 mm) that are not cytotoxic. A special adhesive in conjunction with an
advanced dish design makes the FluoroDish the only non-cytotoxic glass
dish in the world. (See page 210.) This means that cells can actually be
cultured for 4 to 5 days in the same dish and then carry the complete
cell culture with media from the incubator directly into the (Air-Therm
controlled) recording chamber. This is a comfort level that hitherto was
unattainable by scientists and clinical researchers since cells will have
to be transported from the culture dishes on to cover slips just prior to
imagery or attaching patch electrodes.
Chamber: The most problematic aspect of any live cell imaging
procedure or patch clamp approach is the thermal drift of high
magnification lenses caused by temperature fluctuations in the
microscope's environment. It is necessary to constantly readjust the focus
during image acquisition due to this thermal-drift, introducing artifacts
during quantitative microscopy due to the photo-bleaching that occurs
during refocusing. The optimal solution to this problem is to use WPI’s Air-
Therm temperature controller connected to a Plexiglas or acrylic chamber
that encloses the entire imaging area (or at least the objectives and the
stage). Using such a chamber will result in highly stable temperature
control and minimal thermal drift.
Precision Heater Controller
Smart, electrically quiet air heater for live cell
imaging systems and custom incubators
AIR-THERM SPECIFICATIONS
AIR FLOW RATE 20 to 50 CFM (0.55 to 1.4 cubic meter/M)
CONTROL TEMP. RANGE Ambient to 60°C
TEMPERATURE RESOLUTION 0.1 °C
TEMPERATURE ACCURACY 0.2 °C
CONTROL MODES (a) Auto (PID set manually, or Self-Tune)
(b) Manual tune
ANALOG OUTPUT 0.5°C resolution
FOR CHART RECORDER 0-10V, 0 to 100 C (default, adjustable via
controller)
HEATING VOLUME More than 4 cu ft (85 liter), non-circulating
POWER (CE) 300 W, 95-135 V or 220-240 V, 50/60 Hz
DIMENSIONS 61⁄2 x 8 x 71⁄2 in. (15.5 x 21 x 19 cm)
SHIPPING WEIGHT 10 lb (4.5 kg)
Prices shown are in U.S. dollars. Actual charges will vary because of import duty, freight, and currency fluctuations. To obtain an exact quotation, contact your WPI office.
World Precision Instruments •  Tel: 941-371-1003  •  Fax: 941-377-5428  •  E-mail: sales@wpiinc.com  •  Internet: www.wpiinc.com
AIRTHERMY-ATX Air-Therm Heater 120 V, U.S.
AIRTHERMZ-ATX Air-Therm Heater 240 V Specify line cord
2 pieces of 4.5 ft clear, coil-reinforced heater hose included with Air-Therm.
Specify Line (Mains) Voltage
OPTIONAL ACCESSORIES
15590 Clear hose, 2.5” diam., 4.5 ft
300276 Replacement platinum temperature probe
3491 5 ft (1.5m) probe extension cable